- Other Fluke companies:
- Fluke
- Fluke Biomedical
- Fluke Networks
- Fluke Process Instruments
Improved High-Temperature Standard Platinum Resistance Thermometer
To prevent short circuits, to improve stability, and to raise the upper temperature limit to the freezing point of copper (1084.62 ◦C), the high-temperature standard platinum resistance thermometer (HTSPRT) was redesigned. The most important change was an improvement in the structure of the sensor support. The strip support was replaced by a new specially designed cross support. The structure and design of the new HTSPRT are briefly described in this article. The test results of a group of thermometers are presented. The test included long-term drifts of the thermometers at the triple point of water and freezing point of silver during a period of a few hundred hours operation at 1085 ◦C, the short-term stability of R (tpw) and W (Ag) in a period of 5 days, and thermal cycles between 22 ◦C and 1085 ◦C. The test results show that the thermometer performance is improved, and the new HTSPRT can operate up to the freezing point of copper.
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Data Acquisition and Test Equipment
- Temperature Calibration
- Humidity Calibration
- Pressure Calibration
- Flow Calibration
- Process Calibration Tools
- Calibration Software
- Service and Support
- All Calibration Instruments
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us