- Other Fluke companies:
- Fluke
- Fluke Biomedical
- Fluke Networks
- Fluke Process Instruments
PPC4 v1.2.6.14 Embedded Software
TITLE | VER | USE WITH |
PPC4™ | 1.2.6.14 | PPC4 (including -AF models) |
Instructions: | ||
Download the file by clicking on "PPC4 v1.2.6.14.bin".
Run Winflash and select the file downloaded as the file to write to the product. Make sure the correct COM port of the product is connected to a RS232 port on the host PC. Use a null modem cable to COM2 port of PG7000 Piston Gauge. Use a standard RS232 cable to COM1 port of PPC, PPC, RPM and molbox. The updating of embedded software should only be performed after given specific instructions to do so from a Fluke Calibration Authorized Service Center ». If the update isn't performed properly, your product will be inoperable until it is. |
||
Updates Include: | ||
Ver 1.2.6.14 (2020 Mar 07) Fix: Issues related to BG model IL bypass valve not closing to allow normal operation if a vent process is aborted before the vent process is complete Fix: An issue (on PPC4 Basic front panel interface only) where a date greater than 20191231 cannot be entered on the front panel calibration or Time&date menus |
||
Ver 1.2.5.0 (2017 Mar 26) FIX: (PPC-625) Improved control performance for low negative gauge setpoints. |
||
Ver 1.2.4.8 (2016 Jan 11) FIX: (PPC-620) Control may not respond to a new target after 24.5 days of being on. FIX: (PPC-621) PPC4 phase 2 high pressure controller unable to reliably float 200 kPa/kg PC (IP & DP commands not working right) |
||
Ver 1.2.3.38 (2014 Apr 14) ADD: (PPC-614) Increase max number of exercise cycles from 50 to 999999 ADD: (PPC-613) Allow 30 Advanced AutoTests instead of 20 FIX: (PPC-382) Leak test does not work with external RPT FIX: (PPC-386) PPC4 AutoTest does not permit negative gauge mode setpoints FIX: (PPC-389) Watchdog error if IH Q-RPT was last in negative gauge mode and switch to X1H Q-RPT in RPM4 in absolute mode and set pressure lower than ATM FIX: (PPC-393) Decrease sensitivity of knob operations FIX: (PPC-394) PPC4 vent valve opens and then closes during data acquisition at gauge 0 point at the start of COMPASS test FIX: (PPC-396) Absolute mode zeroing with an active external RPM4 QRPT does not work (GUI) FIX: (PPC-397) Autotest not allowing pressure less than full scale FIX: (PPC-405) Control reference setting resets to "Manual Atmosphere" after power cycle FIX: (PPC-410) ARANGE= command will accept negative absolute range requests when RPT to use is explicitly given FIX: (PPC-385) Memory Fault 3.127 FIX: (PPC-409) Cannot run VAC config when in absolute meas mode (Basic UI only) FIX: (PPC-398) mmH2O,20 and mmH2O,60 are incorrect. FIX: (PPC-399) Unit designation for inches and mm of water need to be changed to the current convention of inWa and mmWa from the inH2O and mmH2O. FIX: (PPC-362) Default hold limit minimum for G15k should be 10 ppm of 30 kPa (0.0003 kPa) FIX: (PPC-160) It is not necessary for config to be in absolute or gauge units. It should be able to run in either [GUI]. |
||
Ver 1.2.2.63 (2012 Mar 12) Ver 1.02 (2010 Sep 15) |
||
Ver 1.01a (2010 Jan 22) CHG: Updates a factory setting associated with low absolute pressure control. Affects new builds only. |
||
Ver 1.01 (2009 Dec 21) ADD: Support for “Phase-II” low pressure control when a Phase-II generation module is installed ADD: Support Ethernet as a remote interface option. FIX: “*OPT?” command response missing a comma FIX: A large negative measurement does not generate an under-pressure message on all RPT types. |
||
Ver 1.00d (2009 Jun 15) ADD: Chinese (Simplified), Chinese (Traditional), Czech, English, French, German, Italian, Japanese, Portuguese, Russian and Spanish are supported in the advanced user interface. An Earth icon on the main run screen accesses the language selection screen. FIX: PPC4 with external RPM4 active in negative gauge mode alarms and will not set pressure when indicated pressure is negative. |
||
Ver 1.00c (2009 Mar 02) ADD: Support a "knob jog" secondary pane option when the PPC4 is actively controlling a pressure in dynamic control mode. ADD: Command Interpreter functionality added to interpret the remote commands of other pressure controllers. CHG: Support up to 20 Advanced Auto Tests rather than 10. This firmware release will erase any existing Auto Tests. FIX: AutoTest with non-zero start point still displayed "0 to ..." as label FIX: Cannot edit date for barometer calibration in standard display model. FIX: Change user unit coef range from "0.001,1000" to "0.000001,1000" FIX: PON bit of the "*ESR?" query always clear. |
||
Ver 1.00b (2008 Aug 12) FIX: Low pressure control can be unstable, and not get Ready. FIX: Pressure control should detect instability and make necessary corrections in control times to stabilize. |
- Home
- Products
- New Products
- Electrical Calibration
- RF Calibration
- Data Acquisition and Test Equipment
- Temperature Calibration
- Humidity Calibration
- Pressure Calibration
- Flow Calibration
- Process Calibration Tools
- Calibration Software
- Service and Support
- All Calibration Instruments
- Purchase Info
- News
- Training and Events
- Literature and Education
- Service and Support
- About Us